Labs and lab equipment

Biosafety cabinet after a move

A biosafety cabinet protects the worker, the sample and the environment. Even a small move can change airflow or damage the filter seal.

Why it matters

Shocks in transit can shift the filter or damage its seal. Without testing there's no way to know the cabinet still protects.

What certification includes

  • HEPA filter integrity test
  • Airflow measurement
  • Alarm checks
  • A label with the certification date

How to plan

Book a certified tester for the day after placement. Until then the cabinet isn't used.

About the service: Fume hood moving · All business moving

FAQ

Even a move within the room?

Yes. Any move warrants recertification.

Who certifies?

A certified tester on behalf of the lab.

Do you move the cabinet?

Yes, upright and wrapped, after the lab has cleaned and decontaminated it.